Suggestions/Comments

 

Women's Imaging Clinical Protocols

Routine Female Pelvis without Contrast

Updated 6/12/2001

 

Phased array coil centered over pelvis.

No Gadolinium contrast.

Center on pelvis. Move for HASTE if needed.

Call body radiologist.

Run sequences in the order listed.

Sequence

Plane

Comment

Film #

T2 TSE

Sag

Non-breath hold. 2-3 acq. Small FOV.
Add Cor sat band. Use 4-6 mm slices.

2

T2 TSE

Ax

Orient to uterus.
Non-breath hold. 2-3 acq. Small FOV.
Add Cor sat band.

2

T2 TSE

Obl Cor

Orient to uterus.
Non-breath hold. 2-3 acq. Small FOV.

2

3D T2

Ax

 

0

T1 In/out

Ax

2,1

HASTE

Cor

Thru kidneys. Use body coil if necessary.

2

VIBE

Ax

Coverage from aortic bifurcation to symphysis pubis.

1

 

NOTE: These protocols apply to Siemens Symphony (with Quantum gradients) and Sonata systems. While they reflect the protocols used at NYU Medical Center, NYU is not responsible for their application elsewhere.

 


Department of Radiology
© 2003 NYU School of Medicine
Ethics & Disclaimer