Suggestions/Comments

 

Women's Imaging Clinical Protocols

Routine Female Cervix with Contrast

Updated 12/3/2002

 

Phased array coil centered over pelvis.

Weight based Gadolinium contrast (*), 15cc if pt is less than 180 pounds; otherwise 20cc Gd

Center on pelvis. Move for HASTE if needed.

Call body radiologist.

Run sequences in the order listed.

Sequence

Plane

Comment

Film #

T2 TSE

Sag

Orient to cervix. Non-breath hold. 2-3 acq. Small FOV.
Add Cor sat band.

2

T2 TSE

Ax

Orient to cervix.
Non-breath hold. 2-3 acq. Small FOV.
Add Cor sat band.

2

T2 TSE

Obl Cor

Orient to cervix.
Non-breath hold. 2-3 acq. Small FOV.

2

3D T2

Ax

0

T1 In/Out

Ax

2,1

HASTE

Cor

Thru kidneys.
Reposition patient and use the body coil if necessary.

2

VIBE

Ax

Pre-contrast. Coverage from aortic bifurcation to symphysis pubis.

0

VIBE

Ax

Scan delay = 30 sec.
3 measures (0, 30, 180 sec).

1,2

(*) The use of gadolinium contrast material for these applications represents off-label usage in the U.S. Outside the U.S., please consult your country's regulations for local guidelines.

 

NOTE: These protocols apply to Siemens Symphony (with Quantum gradients) and Sonata systems. While they reflect the protocols used at NYU Medical Center, NYU is not responsible for their application elsewhere.

 