Suggestions/Comments

 

Vascular Imaging Clinical Protocols

Vascular Malformation MRA

Last updated: 1/10/2002

This technique is done to evaluate vascular lesions such as AVMs

40 cc Gadolinium contrast (*) undiluted

If there is an upper extremity lesion, make sure IV is in opposite arm

Use head or extremity coil if possibl

Call Body radiologist.

Sequence

Plane

Comment

Film #

true fisp

Ax

 

0

DB HASTE

Ax

May want to use coronal plane

2

T1

Ax

 

 

T2 TSE

Ax

Fat sat is optional.

 

VIBE

 

 

 

3D FLASH

Cor

With fat sat.  (on 1 or 2, use 3D FLASH found in carotid protocol, change parameters to achieve 6-9 sec. scan time)

0

 

 

NO TIMING RUN

0

3D FLASH

Cor

15-25 measures no gap
Inject gad at 2 cc sec, followed by 20 cc saline at 2cc/sec

0

VIBE

Ax

1 measure through veins of interest.
This can save the case if the venogram doesn’t come out well.

2

SUBS

 

Subtract first few arterial-pre and last 2-3 postcontrast measures-pre.

0

(*) The use of gadolinium contrast material for these applications represents off-label usage in the U.S. Outside the U.S., please consult your country's regulations for local guidelines.

 

NOTE: These protocols apply to Siemens Symphony (with Quantum gradients) and Sonata systems. While they reflect the protocols used at NYU Medical Center, NYU is not responsible for their application elsewhere.

 


Department of Radiology
© 2003 NYU School of Medicine
Ethics & Disclaimer