Suggestions/Comments

 

Vascular Imaging Clinical Protocols

Abdominal Aorta for Mesenteric Ischemia

Last updated: 1/10/2002

Phased array coil centered over abdomen

20 cc Gadolinium contrast (*)

Assess the patient’s breath holding capability. If poor capability, give oxygen. If the patient can’t hold his/her breath call body radiologist.

Run sequences in the order listed.

Sequence

Plane

Comment

Film #

HASTE

Cor and Axial

No fat sat. 5-6mm slices

1

VIBE

Ax

Thru Kidneys

0

3D FLASH

SAG

Use 512 matrix.
Try to get effective thickness less than 1.5 mm.
Want to include celiac and SMA and renal arteries

0

Timing Run

Ax

Thru kidneys – 1cc at 2cc/sec followed by 20 cc saline at  2cc/sec

0

3D FLASH

SAG

Two measures (with 7 sec between)
Second measure may be done on inspiration.
Standard timing run.

0

VIBE

Ax

History to prior

1

Do Subtractions

 

1-Arterial phase – pre-contrast
2 vibe post-pre

0

(*) The use of gadolinium contrast material for these applications represents off-label usage in the U.S. Outside the U.S., please consult your country's regulations for local guidelines.

 

NOTE: These protocols apply to Siemens Symphony (with Quantum gradients) and Sonata systems. While they reflect the protocols used at NYU Medical Center, NYU is not responsible for their application elsewhere.

 


Department of Radiology
© 2003 NYU School of Medicine
Ethics & Disclaimer