Suggestions/Comments

 

Genitourinary Clinical Protocols

Routine Prostate

Updated 6/12/2001

Phased array coil centered over lower pelvis.

No Gadolinium contrast.

Sequence

Plane

Comment

Film #

T1 TSE

Ax

High resolution – 512 matrix. + Mag (small FOV)

2

T2 TSE

Ax

High resolution – 512 matrix. + Mag (small FOV)

2

T2 TSE

Cor

High resolution – 512 matrix. + Mag (small FOV)

2

HASTE

Cor

No fat sat. 5-6mm slices

2

T1 in

Ax

Coverage from aortic bifurcation to symphysis pubis.

2

TrueFISP

Ax

Coverage from aortic bifurcation to symphysis pubis.

2

 

NOTE: These protocols apply to Siemens Symphony (with Quantum gradients) and Sonata systems. While they reflect the protocols used at NYU Medical Center, NYU is not responsible for their application elsewhere.

 


Department of Radiology
© 2003 NYU School of Medicine
Ethics & Disclaimer