Suggestions/Comments

 

Genitourinary Clinical Protocols

Routine Bladder

Updated 6/12/2001

Phased array coil centered over pelvis.

Weight based Gadolinium contrast (*), 15cc if pt is less than 180 pounds; otherwise 20cc Gd

Run sequences in the order listed.

Sequence

Plane

Comment

Film #

T2 TSE

Sag

Non-breath-hold. (from female pelvis)  (4mm 0.2 gap)

2

T2 TSE

Ax

Non-breath-hold. (from female pelvis)  (4mm 0.2 gap)

2

T1 in

Ax

Coverage, bifurcation to pubic symphysis.

2

HASTE

Cor

Thru kidneys. Use body coil if necessary.

2

VIBE

Ax

Small FOV.
Try for 2 mm thickness.

2

VIBE

Ax

Scan delay = 45 sec.
2 measures (0 and 180 sec).

1,2

(*) The use of gadolinium contrast material for these applications represents off-label usage in the U.S. Outside the U.S., please consult your country's regulations for local guidelines.

 

NOTE: These protocols apply to Siemens Symphony (with Quantum gradients) and Sonata systems. While they reflect the protocols used at NYU Medical Center, NYU is not responsible for their application elsewhere.

 


Department of Radiology
© 2003 NYU School of Medicine
Ethics & Disclaimer